Flow-Through Cells with Oxygen Sensor

  • Flow-through cells with integrated contactless optical oxygen sensor
  • Luer-lock with tubing adapters
  • Works with gas (%O2) and with liquids (DO)
  • No oxygen consumption
  • Negligible drift
  • >3 years shelf time
  • Trace oxygen versions available

The flow through cells have an integrated oxygen sensor spot, allowing precise contactless oxygen measurements of gaseous or liquid samples flowing through the cell. The flow-through cells are available as a small version OXFTC and as a large version OXFTC2. The cell is equipped with luer-lock connectors on both tubing ends. A set of luer-lock adapters is included allowing direct connection of tubings with ca. 1-2 mm inner diameter (OXFTC) or ca. 3-4 mm inner diameter (OXFTC2). The flow through cell is connected via an optical fiber to an optical oxygen meter from Pyro Science (see accessories below).

Trace range versions. A trace range version TROXFTC is available for high precision measurements close to 0% O2. It offers about 5x higher resolution compared to the normal range version. The recommended measuring range is 0-10 %O2. However, calibration at ambient air (21% O2) is possible, ensuring a straightforward calibration without the need of special calibration standards.

Gallery

Compatible oxygen meters

FireStingO2 (FSO2-x)
FireStingGO2 (FSGO2)
FireStingO2-Mini (FSO2-MINI-ST)
Piccolo2 (PICO2)

Available Items

Item No. Description Response
Time (s)**
Pricing
in EU*
OXFTC Flow-Through Cell with optical oxygen sensor (2 pieces per package), incl. Luer-lock adapters for ca. 1-2 mm ID tubings, recommended flow rate: 10-100 mL/min <3s (liquid)
<1s (gas)
200 €
OXFTC2 Flow-Through Cell with optical oxygen sensor (2 pieces per package), incl. Luer-lock adapters for ca. 3-4 mm ID tubings, recommended flow rate: 20-500 mL/min <3s (liquid)
<1s (gas)
200 €
TROXFTC Flow-Through Cell with trace oxygen sensor (2 pieces per package), incl. Luer-lock adapters for ca. 1-2 mm ID tubings, recommended flow rate: 10-100 mL/min <5s (liquid)
<1s (gas)
200 €

Accessories for Connection to FireSting Devices

Item No. Description Pricing
in EU*
SPFIB-BARE Bare Optical Fiber (1m length), ST-plug + one polished end 50 €
SPFIB-BARE-CL2 Bare Optical Fiber (2m length), ST-plug + one polished end 60 €

Accessories for Connection to Piccolo2 Devices

Item No. Description Pricing
in EU*
PICFIB2 Optical Fiber (1m length) for Piccolo2, bare polished end 70 €

* Only EU-Prices shown. Contact us for pricing outside EU. Not for resale.

** Typical response times for 90% signal change (T90).
     For liquids: measured for the transition from air into a stirred solution of 1% Na2SO3.

Specifications

Given for 2-point calibrated oxygen sensors at 1013mbar absolute pressure.

Normal range versions hPa %O2 %air sat. mg/L
Optimum measuring range 0-500 0-50 0-250 0-23
Maximum measuring range (not specified) 0-1000 0-100 0-500 0-45
Detection limit 0.2 0.02 0.1 0.01
Resolution at 1% O2 0.1 0.01 0.05 0.005
Resolution at 20% O2 0.5 0.05 0.25 0.025
Accuracy at 1% O2 ±0.2 ±0.02 ±0.1 ±0.01
Accuracy at 20% O2 ±2 ±0.2 ±1 ±0.1
 
Trace range versions hPa %O2 %air sat. mg/L
Optimum measuring range 0-100 0-10 0-50 0-5
Maximum measuring range (not specified) 0-210 0-21 0-100 0-9
Detection limit 0.05 0.005 0.02 0.002
Resolution at 0.2% O2 0.02 0.002 0.01 0.001
Accuracy at 0.2% O2 typ. 2% of reading, but not better than 0.01%O2
All versions
Temperature range 0 to 50°C (specified), -20 to 70°C (not specified)
Minimum lifetime 10,000,000 data points
Application areas Laboratory, industry, research. Not for medical or any other safety-critical applications. Not for application in humans or in food intended for human consumption.
Compatibility Gas-phase, aqueous solutions, ethanol, methanol, isopropanol
Cross sensitivity Other organic solvents, chlorine
No cross sensitivity pH 1-14, CH4, CO2, H2S, any ionic species
Sterilization Ethylene oxide (EtO), 70% ethanol, 70% isopropanol
Cleaning 3 % H2O2, ethanol, soap solution
Storage time >3 years in darkness at room temperature
Calibration Single point, two point calibration
Optical isolation The sensor spots are covered with a final black layer in order to minimize influence of strong external illumination
Tubing connector Luer-lock with adapters for ca 1-2 mm ID tubings (OXFTC) or 3-4 mm ID tubings (OXFTC2)